Monday, January 5, 2009 TCU: Directions for UseI. INTRODUCTIONIf you need assistance related to this product, please call Geiger Medical Technologies at 800.771.9577 (949.240.7584). It is important that the instructions supplied with this equipment be read, understood, and followed in order to enhance safety and effectiveness. II. TABLE OF CONTENTSIII. PRECAUTIONSCAUTION: Federal (U.S.A.) law restricts the sale of this device to or on the order of a physician or healther care profession. CAUTION: This device is supplied non-sterile. WARNING: The risk of ingniting flammable gases or other materials is inherent in the practice of thermal cautery and cannot be eliminated by device design. Precautions must be taken to avoid contact of flammable materials and substances with thermal cautery electrodes, whether they are in the form of an anesthetic or skin preparation agent, are produced by natural processes within body cavities, or originate in surgical drapes, tracheal tubes, or other materials. CAUTION: Studies have shown that smoke generated during thermal cautery procedures may be harmful to surgical personnel. These studies recommend using a surgical mask and adequate ventilation of the smoke using a smoke evacuator or other means. CAUTION: The Geiger TCU handpiece is designed for intermittent use only. Continuous activation is not recommended and will cause permanent damage to the handpiece. CAUTION: The cautery electrode shafts, insulation material, and the handpiece metal nosecone may become hot with use. Do not allow these areas during or immediately after use to make human contact or contact with materials that may be adversely affected by heat. IV. INDICATIONS FOR USEThe Geiger TCU is a non-sterile, reusable, AC-powered thermal cautery device. The TCU is intended to be used for destroying small lesions and to coagulate small bleeders by applying high levels of heat via a cautery electrode. V. INSPECTIONThe Geiger TCU includes: One (1) Thermal Cautery Unit Part No.: 150 One (1) Cautery Handpiece Part No.: 250 Three (3) General Purpose Cautery Electrodes Part No.: 201 A, C, & E Operation Manual Part No.: 81-0018 Upon receipt of your Geiger TCU, carefully unpack the contents and verify that all items ordered are present and undamaged. If damage has occurred during shipment, or an item is missing, contact your Geiger Distributor or Geiger directly within ten days of purchase. To purchase additional TCU supplies, contact your Geiger Medical Technologies Distributor, or contact Geiger Medical Technologies at 800.771.9577 (949.240.7584). VI. DEVICE DESCRIPTIONPlease refer to Figure 1 for the following device descriptions:
CAUTION: To reduce the risk of an electric shock, do not remove the rear cover of the instrument. Refer all servicing to qualified personnel.
Figure 1
VII. OPERATING INSTRUCTIONSCAUTION: If you are unfamiliar with the practice of thermal cautery, it would be wise to practice at the various settings on a piece of raw meat. CAUTION: If using the TCU in a horizontal position, leave the handpiece
If the output of the unit is intermittent, make sure that the cautery electrode is making satisfactory contact with the walls of the handpiece receptacle holes. You may want to slightly bend the electrode shafts to increase the contact area if the electrode is loose. CAUTION: The Geiger TCU is intended for use only with Geiger Cautery Electrodes. CAUTION: Over-heating the Cautery Electrode will shorten the life span of the Electrode. CAUTION: When unsure of the output level to use during a procedure, begin at a low setting and increase the output gradually to prevent unnecessary damage to the patient. CAUTION: Disconnect the unit power cord fromt he power receptacle prior to performing cleaning operations. VIII. UNIT CLEANINGCAUTION: To reduce the risk of an electric shock, do not remove the rear cover of the instrument. Refer all servicing to qualified personnel. Periodic external cleaning of the Geiger TCU is required. Clean all external surfaces with a mild soap and water solution using a moist, soft, lint-free cloth. Dry all surfaces with a clean, lint-free cloth. CAUTION: Care must be exercised to ensure that the electrical connections IX. STORAGEStore the Geiger TCU in a cool, dry place. X. STERILIZATION OF ACCESSORIESThe Cautery Handpiece and Cautery Electrode may be autoclaved according to the instructions below. Separate the electrode from the Handpiece prior to autoclaving and use aseptic procedures when handling and sterilizing both items.
CAUTION: Do not use liquid sterilizing solutions with the Geiger TCU or Geiger has exercised care in the manufacturing of the TCU. The warranty period for the unit is one year from the date of purchase and covers parts and labor. The warranty for the Handpiece and Electrodes is 90 days. To initiate and register your warranty, please register your product by emailing us with the item and date purchased along with your name, address, phone, fax and email address. Items for return should be sent with all original packaging and instructions included. Please use proper packaging to prevent damage during shipment. We recommend that you insure the return with your chosen carrier for loss or damage. Include a copy of the original invoice with the return and state the reason for the return. Freight charges to return merchandise to Geiger are non-refundable except in the case of shipping error on the part of Geiger. Please contact Geiger at 800.771.9577 (712.323.3432) to advise that you are returning a Geiger product. XII. SPECIFICATIONS
XIII. SYMBOL EXPLANATIONThe symbols that appear on the TCU, along with their explanations. XIV. VOLTAGE/CURRENT VS. DIAL SETTINGThe following chart contains the voltage and current output of the TCU over the range of dial settings available.
|
|||||||||||||||||||||||||||||||||||
|
Geiger Medical Technologies 608 13th Ave. Council Bluffs, IA USA 51501-6401 voice: (800) 771-9577 (712) 323-3432 fax: (800) 320-9612 (712) 323-1156 info@geigermedical.com |
|||||||||||||||||||||||||||||||||||